Measurement methods | Potassium dichromate digestion at high temperature, colorimetric determination (National Standard GB11914-89) |
Test range | (0-100), (0-200), (0-1000), (0-5000) mg / L range is optional and expandable |
Maximum test range | 0-40000mg / L (after automatic dilution) |
Resistance to chloride ion interference | Maximum 5000mg / L (after automatic dilution) |
Detect offline | 8mg/L |
Resolution | <1mg/L |
Accuracy | Standard solution <5%, water sample <10% |
Reproducibility | <5% |
Digestion temperature | 165℃,can be set |
Digestion time | 15min,can be set |
Trouble-free running time | ≥720h/ time |
Range drift | ±5%F.S |
Sample interval | Continuous, 1 hour, 2 hours, 24 hours, trigger |
Calibration interval | Manually or automatically at selected intervals and times (1-7 days) |
Reagent consumption | About 720 samples per set of reagents |
HMI | 7 inch, 70,000 colors, 800 * 480 resolution, TFT true color touch screen |
Print | Reserved printer interface, can be connected to industrial micro printer (optional) |
Store | 20,000 pieces of data, which will not be lost when power is off |
Communication Interface | 1 channel RS232 digital interface or RS485, support MODBUS communication protocol or custom protocol 1 channel analog 4-20mA (20mA corresponding range adjustable) |
Pretreatment system | The functions of self-cleaning, back-blowing, and precision filtering ensure that the sample is well represented, and also avoid the suspension of suspended particles in the pipeline to ensure the continuity of data. (Optional) |
Dimensions | 770*600*450(mm) | Weight | 50kg |
Power supply | AC220V±20%,50Hz±1% | Power | 300W |
Ambient temperature | 5—40℃ | Environment humidity | ≤85% |